<span id="lkot54ftt"><caption id="blu3jlo3" class="nvsvc7g"><s id="ed1n0sy7wp3" class="vdbxvt072zf"></s></caption></span>
    Hi!Welcome to Taian Demei Electromechanical Equipment Co.,Ltd.!
    Taian Demei Electromechanical Equipment Co.,Ltd.

    Focus on the production of temperature calibration experimental equipment field for ten years

    Research and development Center of thermal measurement calibration experimental equipment national epidemic prevention unit

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator

    ◇ Product Introduction ◇ Product Features◇ Technical Indicator

    [Product Introduction]

    DY-RX02 Thermal Resistance Calibration Device is a hand-held tool used to calibrate RTD (Temperature Measuring Resistance Body) temperature transmitters. It can simulate and measure many different types of RTDs, and it can also simulate and measure resistors.

    [Product Features]

    1. Strong protection function: waterproof level IP67, any signal end misconnect 220V automatic protection;

    2. Measure the output temperature of the thermal resistance;

    3. Simulated thermal resistance output;

    4. Accuracy: DY-RX02A 0.02%, DY-RX02B 0.05%;

    5. Provide manual stepping, automatic stepping, 0 ~ 100% step and slope output function;

    6. With white LED backlight, manual backlight adjustment and automatic power off function, suitable for field use;

    7. Unique automatic identification of three-wire, four-wire wiring mode;

    8. Battery box cover, convenient battery replacement;

    [Technical indicators]

    General Technical Specifications

    1. Operating temperature: -10℃ ~ 55℃

    2. Storage temperature: -20℃ ~ 60℃

    3. Relative temperature: 0 ~ 90%RH, non-condensation

    4. Weight: 350g

    5. Dimensions: 47 mm×93 mm×185 mm

    6. Power supply: 6 7# batteries or AC adapter

    Standard accessories

    1. Watch pen: 18A53 watch pen 1 pay

    2. Power Adapter: One 18A52 charger

    DY-RX02 Thermal Resistance Calibration Device

    List of output and measurement functions

    FunctionMeasurementOutput
    Direct Voltage0~30VNone
    Resistance0~3200Ω0~3200Ω
    Thermal ResistancePt100、Pt1000、Cu50、Cu100Pt100、Pt1000、Cu50、Cu100
    OtherStep output, slope output, step output, custom range

    Measurement Index

    FunctionMeasurementAccuracy
    DY-RX02ADY-RX02B
    Voltage0~30.000V0.02%+20.05%+2
    Resistance(2 wire, 3 wire)0.00Ω~440.00Ω0.15Ω0.25Ω
    400.0Ω~3200.0Ω1.0Ω1.5Ω
    Resistance(4 wire)0.00Ω~440.00Ω0.1Ω0.15Ω
    400.0Ω~3200.0Ω0.5Ω1.0Ω
    Thermal ResistancePt100、Pt1000、Cu50、Cu100//

    Unless otherwise stated, all indicators are based on a one-year calibration cycle and are applicable to temperatures ranging from +18 ° C to +28 ° C. All indicators assume a 10-minute warm-up time.

    Output Indicators

     FunctionMeasurementAccuracy
    DY-RX02ADY-RX02B
    Resistance1.00Ω~440.00Ω0.15Ω0.25Ω
    400.0Ω~3200.0Ω0.51.0
    Thermal ResistancePt100、Pt1000、Cu50、Cu100//

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator

    DY-RX02 Thermal Resistance Calibrator 2020-12-9 This article was read 3294 Times
    Previous:PartnerNext:Partner

    Technical

    Contact us

    Taian Demei Electromechanical Equipment Co.,LTD.

    Address:No. 8, Hedong Road, Shengzhuang Town, Taishan District, Taian, Shandong,China (Mainland)

    Tel: +86-538-6518016 / 6518018

    Mobile: +86-186 5382 5886

    Whatsapp: +86-186 5382 5886   Skype: +86 186 5382 5886

    Contact: Grace   Email: grace@demeithermo.com

    Sales network

    Taian Demei Electromechanical Equipment Co.,LTD.                          Publicity